Page 1
Measurement made easy Introduction — Navigator 540 transmitter The Navigator 500 range of analyzers from ABB are The following parameters are available in the designed for high purity water treatment Navigator 500 range: Navigator 500 applications and power cycle chemistry monitoring.
Page 2
For more information Further publications for the Navigator 500 series are available for free download from www.abb.com/analytical or by scanning these codes: Navigator ADS550 Navigator AHM550 Navigator ASO550 Dissolved oxygen Hydrazine analyzer Sodium analyzer analyzer Search for or click on...
The signal word 'IMPORTANT Electrical Equipment for Measurement, Control and Laboratory (NOTE)' does not indicate a dangerous or harmful Use' and complies with US NEC 500, NIST and OSHA. situation. COM/AWT540/PB–EN Rev. C | Navigator 540 | Transmitter...
This symbol, when noted on a product, indicates a ABB is committed to ensuring that the risk of any potential hazard which could cause serious personal environmental damage or pollution caused by any of injury and / or death.
2.3.2 PROFIBUS and ABB products Navigator AWT540 utilizes PROFIBUS DP as this protocol is optimized for high speed and low connection costs (see www.abb.com/fieldbus and follow the PROFIBUS link). COM/AWT540/PB–EN Rev. C | Navigator 540 | Transmitter...
Read Only Back Select Calibrate Advanced Ethernet – refer to Section 6, page 15. Service Back Select 3. Use the key to scroll to Advanced and press the key (below the Select prompt). Navigator 540 | Transmitter | COM/AWT540/PB–EN Rev. C...
D-type connectors. Positive (+) data line: connect to pin 3 on Suitable PROFIBUS cable (part numbers PCA010, PCA011 and PROFIBUS DP equipment with 9-way PCA012) can be obtained from ABB. Refer to Data Sheet D-type connectors. 10/63-6.46 EN. Braided N / A Cable screen –...
(below the Select prompt). when configuring a PROFIBUS bus system. IMPORTANT (NOTE) The Baud Rate is a read-only value. The current Slave Address setting (for example 70) is displayed. COM/AWT540/PB–EN Rev. C | Navigator 540 | Transmitter...
The ABB GSD file (Ident No. 3402) is divided into 2 sections: 4.7 Navigator AWT540-DP PROFIBUS datasheet — General specifications Item Detail PROFIBUS device NAVIGATOR 500 Identification of the device, together with hardware and name (product: Navigator AWT540) software versions, baud rates supported and the possible...
— DPV0 and DPV1 protocol services. — Navigator AWT540-DP device-specific GSD file (ABB_3402.gsd). A copy of the conformance certificate is available for download from the Navigator AWT540 product pages of the ABB website www.abb.com. COM/AWT540/PB–EN Rev. C | Navigator 540 | Transmitter...
WARNING – Bodily injury Tx Delay Set a delay to the response from the 50 ms Refer to the Navigator AWT540 Commissioning transmitter in milliseconds. Maximum instruction (CI/AWT540-EN) before making electrical delay 100 ms. connections. Navigator 540 | Transmitter | COM/AWT540/PB–EN Rev. C...
— Up to 300 m (984 ft.) – twin twisted pair with overall foil screen and an integral drain wire. — Up to 1200 m (3937 ft.) – twin twisted pair with separate foil screens and integral drain wires. COM/AWT540/PB–EN Rev. C | Navigator 540 | Transmitter...
(the email can be sent to up to 3 recipients). Invert 1 (to 4) If enabled, an email is generated when the Source becomes inactive instead of active. Navigator 540 | Transmitter | COM/AWT540/PB–EN Rev. C...
Inactive In manual test Simple Unsigned 8 Inactive In configuration Simple Unsigned 8 Inactive Media card full Simple Unsigned 8 Inactive Media near full Simple Unsigned 8 Inactive Table A.2: Transmitter diagnostics COM/AWT540/PB–EN Rev. C | Navigator 540 | Transmitter...
Page 22
Sensor 3 – Temperature range high Simple Float Sensor 3 – Temperature range low Simple Float Sensor 4 – Temperature range high Simple Float Sensor 4 – Temperature range low Simple Float Table A.12: Measurement range limits COM/AWT540/PB–EN Rev. C | Navigator 540 | Transmitter...
Page 27
Analog output 2 – Simple Float 22 mA Failure current Analog output 3 – Simple Float 22 mA Failure current Analog output 4 – Simple Float 22 mA Failure current Table A.22: Analog output settings Navigator 540 | Transmitter | COM/AWT540/PB–EN Rev. C...
Page 38
0 and See source list Source 9 to 64 Digital I/O 6 – Simple Unsigned 8 0 to 1 0 – Inverted Polarity inverted 1 – Non inverted Table A.32: Digital I/O settings COM/AWT540/PB–EN Rev. C | Navigator 540 | Transmitter...
Page 42
Table B.5: Sensor signals view 30172 Analog output % of full scale AOUT2 30174 Analog output % of full scale AOUT3 30176 Analog output % of full scale AOUT4 Table B.7: Analog outputs COM/AWT540/PB–EN Rev. C | Navigator 540 | Transmitter...
Sensor 1 – Hardware version number 32001 Sensor 2 – Hardware version number 32002 Sensor 3 – Hardware version number 32003 Sensor 4 – Hardware version number Table B.12: Sensor hardware version (valid range 1 to 9) Navigator 540 | Transmitter | COM/AWT540/PB–EN Rev. C...
Analog output 1 – Failure current 0 – 22 mA 40082 Analog output 2 – Failure current 40084 Analog output 3 – Failure current 40086 Analog output 4 – Failure current Table B.18: Analog output settings Navigator 540 | Transmitter | COM/AWT540/PB–EN Rev. C...
Page 49
1 – Two months 2 – Three months 3 – Four months 4 – Six months 5 – Twelve months Table B.22: Multi-stream sensors – units, filter and scheduled calibration configuration (continued) Navigator 540 | Transmitter | COM/AWT540/PB–EN Rev. C...
Page 50
Analog output 4 – Output type See Analog output 1 – Output type 43120 Analog output 4 – failure current 0 – Enabled enable 1 – Disabled Table B.27: Analog output settings COM/AWT540/PB–EN Rev. C | Navigator 540 | Transmitter...
Sensor 4 function check Transmitter failure Transmitter out of specification Transmitter maintenance Transmitter function check Sensor 1 calibration in progress Sensor 2 calibration in progress Sensor 3 calibration in progress Table C.3: Digital signal sources COM/AWT540/PB–EN Rev. C | Navigator 540 | Transmitter...
Page 53
Acknowledgements PROFIBUS is a registered trademark of PROFIBUS and PROFINET International (PI). Modbus is a registered trademark of the Modbus-IDA organization.
Page 54
We reserve the right to make technical changes or modify the contents of this document without prior notice. With regard to purchase orders, the agreed particulars shall prevail. ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document.