Page 1
HP 3324A Synthesized Function/Sweep Generator Operating and Programming Manual HEWLETT PACKARD...
Page 3
Operating and Programming Manual HP 3324A Synthesized Function/Sweep Generator SERIAL NUMBERS This manual applies to instruments with serial number 2836GOOlll and higher. Any change made in instruments having serial numbers higher than the above numbers will be found in a “Manual Changes”...
Page 4
Notice Subject Matter Notice The information in this document is subject to change without notice. Hewlett-Packard makes no warranty of any kind with regard to this printed material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Hewlett-Packard shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of...
Page 5
Printing History New editions are complete revisions of the manual. Update packages, which are issued between editions, contain additional and replacement pages to be merged into the manual by the customer. The date on the title page and back cover of the manual only changes when a new edition is published.
Page 6
Preface Introduction This manual provides the information required when either operating the HP 3324A from the front panel or programming it from a remote controller. What is in this Chapter 1: Instrument Introduction Gives a brief description of the manual? instrument and its features.
Page 7
Chapter 12: Performance Tests Gives the procedures to verify the in-spec performance of the BP 3324A. Specifications Appendix A: Lists the HP 3324A specifications. Appendix B: Options and Accessories Gives details of all the options and accessories available for the HP 3324A.
Page 8
During the warranty period, HP will, at its option, either repair or replace products which prove to be defective. For warranty service or repair, this product must be returned to a service facility designated by HP.
Page 9
The remedies provided herein are Buyer’s sole and Exclusive Remedies exclusive remedies. HP shall not be liable for any direct, indirect, special, incidental, or consequential damages, whether based on contract, tort, or any other legal theory. Assistance Product maintenance agreements and other customer assistance agreements are available for Hewlett-Packard Hewlett-Packard Sales and Service Office.
Contents Instrument Introduction Introduction ... . . l - l Features ... . . l - l Operating Concept ..Local Mode .
Page 12
HP-IB Connector (16) ..Main Signal Output Connector (17) . . Phase Cal In (18) ..Phase Cal Out (19) ..
Page 13
... . . Powering Up ... . . The HP 3324A as a Function Generator . The HP 3324A as a Sweep Generator . . 5. Parameters Introduction .
Page 14
Setting the Default Parameters ..Choosing the Number of Stores/Intervals Changing the HP-IB Address ..Choosing HP 3325A Compatibility . . . Performing a Self- test ..
Page 15
Omitting the Mnemonic ..9-11 . . . Synchronizing after a Syntax Error 9-11 Setting the HP-IB Address ..Status Reporting Introduction ... . . Status Byte .
Page 16
Performance Tests 12-1 Introduction ... . . 12-1 Equipment Required ..12-1 Test Record ... . . 12-2 Test Failure .
Instrument Introduction The III’ 3324A Synthesized Function/Sweep Generator Introduction a multi-task generator, which can be used as: a reference source Produces a sinewave of a specified frcqucncy, amplitude, DC offset and phase. a function generator Produces various waveforms at a specified frequency, amplitude, DC offset and phase.
Operating Concept The HP 3324A has been designed so that it can be used in either of two modes: a. Local Mode: Using the front panel to initiate operations. b. Remote Mode: Using an external computer as the controller. Local Mode...
About this Manual This manual is basically divided into two main parts: A part which describes the local control of the instrument (chapters 2 to 8). A part which describes the remote control of the instrument (chapters 9 to 11). The examples in the local control part are shown in such a way that on the left side of the page the keystrokes are shown, while the right side gives an explanation of...
About the Note The NOTE sign is used to show where additional Note information beneficial to the reader can be found. This information refers to specific situations or operations. Instrument Introduction...
Front and Rear Panel Overview Introduction This chapter contains basic information about the front and rear panels of the HP 3324A. The descriptions given should be read in conjunction with the foldouts at the front of this chapter. Front Panel The front panel can be divided into 15 main component areas, each of which is described here.
SELECT/MODIFY (4) These keys are used to alter data and to move the cursor within menus or the display. The function of the keys depends on the current menu or value displayed. For more information see Chapter 3 Instrument Operation. The PARAMeter keys allow you to view, enter or change the frequency, amplitude, DC offset and phase of the...
If (Shift is pressed “S” is shown in the display until the next key is pressed. This key returns the HP 3324A from remote control to front-panel control unless local lockout has been programmed. The Iut;l) key accesses the Utilities menu, which contains...
The rear panel can be divided into 13 main component areas, each of which are described here. HP-IB Connector (16) The HP-IB connector is used to connect the HP 3324A onto the Hewlett-Packard Interface Bus so that it can be used in remote mode and accessed by a controller.
30 MHz oscillator. If an external reference is present “EXT-REF” is shown in the display. Ref Out (23) A 1 MIIz signal from the IIP 3324A reference circuits is available here. Aux 0 dBm (24) A squarewave signal is available at this output for frequencies between 21 MIIz and 60 MHz.
Instrument Operation Introduction This chapter describes the operating principles of the HP 3324R. It describes the powering on procedure, the various outputs available and how to use both the DATA keys and the SELECT/MODIFY keys. Before Power-on Caution Before connecting the HP 3324A...
HP Service Center for repair. Turn-on State The initial state of the HP 3324A depends on if the instrument is in the HP 3325A compatibility mode or not. The mode can be set and seen in the Utilities menu (see Chapter 8).
Page 37
In the HP 3325A compatible mode all stores are deleted, and a default setting becomes the new state. Table 3-l shows the power-on state for the HP 3325A compatible mode. The main signal output is enabled at power-on (LED to the right of the (SignalG&ZKX) key is lit).
Input/Output The HP 3324A has a total of 13 BNC connectors, located on both the front and rear panels. The use of Signals each connector is explained in the following sections. Caution A maximum of 42 V can be applied as an external voltage to any of BNC connectors because of the HP 3324A’s floating ground.
Phase Calibration This input must be connected to the Phase Cal Output of another HP 3324A. One of the instruments must Input have option 003 installed and the other option 004. The two instruments can then be phase calibrated. Use the...
Reference Input oscillators). Phase-locking to an external frequency reference transfers the external reference’s frequency accuracy and aging rate to the HP 3324A. The external reference controls the standard 30 MHz internal reference oscillator frequency. The external reference level must be from 0 dBm to +20 dBm (50R), and the frequency...
The Ref Out 1 MHz connector supplies a 1 MHz squarewave derived from the frequency reference of the output HP 3324A. The squarewave has a level greater than 0 an can be used to phase-lock an analyzer or other instrumentation to the frequency reference of the HP 3324A.
X-drive The X-drive output signal is only present during a sweep operation, as follows: Single sweep: 0 V at the start of the sweep, increasing linearily to > +lO V at the sweep stop. The voltage remains at this level until the reset prior the start of another sweep.
Aux 0 dBm 21-60 MHz The rear-panel Aux 0 dBm 21 - 60 MHz connector supplies a squarewave signal for frequencies between 21 and 60 MHz. The amplitude of the signal is 0 dBm. To obtain a 21 - 60 MHz frequency, the auxiliary TTL clock waveform (“AUX”) must be selected.
As on calculators the (BackSpace) key clears the digit to the left of the cursor. If an entered value exceeds the range limits, the HP 3324A ignores the entered value, briefly displays the error, and then puts the instrument into the edit mode.
SELECT/MODIFY The SELECT/MODIFY keys are used in conjunction with most of the other keys and the display. The four Keys arrow keys are referred as the a, a, a and a keys in this manual. Select The select mode is used to move through the menus to a certain item/point, which can then be activated with the (Select) key.
Page 46
There are two ways to change a digit in the modify mode: the numeric keypad. There is no need to press [Enter). Use the Ilncr)/m keys to increase/decrease the digit by 1. Holding the key in will increase/decrease the value at a faster rate. Note that, except on the most significant digit, increasing from 9 to 10 causes the 1 to be carried to the next digit position, and decreasing from 0 to 9 will cause the preceeding digit to be...
Page 47
Example Consider a frequency of 125 KHz. You wish to see what happens when you increase the frequency to 225 KHz and back again in steps of 10 KHz. Result The current frequency value is displayed is set. A frequency of 125 KHz entered modify mode is and the cursor is placed under...
Example Consider an amplitude of 4.075 V(p-p). You wish to see what happens when you increase the amplitude to 4.125 Keystrokes The current amplitude is shown in the display. The amplitude is changed to 4.075 V(p-p). The edit mode is activated. The cursor moves under the “5”.
(St,: xx int.: XX) command contained in the Utilities menu (see Chapter 8). At power-on, if the HP 3324A is in the HP 3325A compatible mode, all the stores will be deleted. If the instrument is in the non-compatible mode, the stores will contain the same information as previous to power-off.
Example Consider saving an instrument state in the 4th register. Keystrokes The present instrument setting is saved in register 4. operating Recalling Instrument state is recalled from the internal memory, when the m key, a digit and then (Enter) are pressed. States digit specifies the memory location from which the operating state is to be recalled.
Introduction This chapter gives a very brief description of how to get started with the HP 3324A. This includes how to power up, an example of setting the parameters for a standard function generator, and an example of setting up a sweep.
(LED to the right of the I-j switch is lit). The HP 3324A as The following is an example of how to set up the HP 3324A as a standard function generator. It shows how a Function to choose a waveform, set the frequency, amplitude and Generator offset of the output signal, and how to enable the signal.
Page 53
Keystrokes Result The triangle is selected as the signal waveform Note If the error “freq. out range” is shown here, do steps 3 and 4 before repeating steps 1 and 2. This is because the frequency is not valid for the waveform chosen. Keystrokes Result The parameter menu is shown in the display, along with the...
Page 54
Keystrokes Result The LED to the left, of the key lights, to show that the main signal output is enabled. This example has shown how to set up the HP 3324A as a function generator. 4-4 Getting Started...
The HP 3324A as a The following is an example of how to set up the 3324A as a sweep generator. The example follows Sweep Generator on from the one above. It shows how to set up three sweep intervals, and to cause a continuous sweep to be performed using these intervals.
Page 56
Note The sequence does not have to follow in series, that is 1, 2, 3, 1, 2. You can program the intervals in any sequence that you require, for example, 2, 3, 1, 3, 1, 2 and so on. Now, each of the intervals must be programmed separately.
Page 57
Keystrokes The stop frequency for the 2nd interval is changed to 10 1 5 . The sweep time for the 2nd interval is changed to 2 s. The cursor is now on “SWEEP-TIME”. The 2nd interval is set as a logarithmic sweep. The cursor is now on “LOG”...
Parameters Introduction The Parameter keys (Freg), enable you to display the frequency, amplitude, DC offset and phase values, respectively, of your output signal. Figure 5-1. The Parameter Keys When one of the keys is pressed, the current value is displayed along with the parameter key functions and the current waveform.
K H z RMPL OFFS Figure 5-2. The Parameter Display One of the other parameters can be viewed by either pressing the parameter key, or using the 0J or a keys to move the cursor up or down the Parameter menu. Putting the cursor (so that the parameter blinks) on the amplitude, on “OFFS”...
Modifying A parameter can be simultaneously updated using the modify keys. This is done by activating the parameter Parameter that you wish to change, and then pressing the yellow “M” is shown in the display, to indicate that you are in the modify mode.
Modify Example Consider a frequency of 125 KHz. You wish to see what happens when you increase the frequency to 225 KHz and back again in steps of 10 KHz. Note Make sure that the waveform is a sinewave or squarewave before doing this example, otherwise the frequency set will be too invalid and an error will be caused.
There are two ways to change a digit in the edit mode: Type in a new digit value by pressing the number on the numeric keypad. There is no need to press IEnter). Delete the digit before the cursor by pressing the key.
Press IFreq) (frequency) to display the frequency of the Frequency output signal. To change this value either enter a new value with the numeric keypad, and press the relevent unit key; LMHz), ml, or @. edit the existing value using the edit function. Resolution of the frequency entry is 1 mHz for frequencies upto 999,999.999 Hz, independent of the waveform selected, and from 1 MHz upwards the...
Example The following example shows how to change the frequency to 10.7854 KHz:- Result The Parameter display is shown. “FREQ” is high-lighted. “10.7854” is displayed. The frequency is then accepted and “10.7854 KHz” is displayed. Amplitude Press IAmpl (amplitude) to display the amplitude of the output signal.
Amplitude Calibration An amplitude calibration of the output signal can be performed on the HP 3324A. To do this, press the (Ampl) key (m a). The main signal output is switched off while the calibration is in progress.
High-Voltage Option When the high-voltage output is used (option 002 is installed), a maximum output of 40 V peak-to-peak is available into a high impedance. The load resistance must be more than 50052 or distortion will result, particularly at higher frequencies. To assure square wave overshoot of <...
Example The following example shows how to change the amplitude to 2.309 Vrms (6.531 V(p-p)) Keystrokes Result The Parameter display is shown. “AMPL” is highlighted. “2.309” is displayed. The amplitude is accepted and “2.309 V RMS” is displayed. DC Offset The m (DC offset) key enables you to display the DC offset of the output signal.
Offset Only, No AC When the DC function is activated in the waveform menu (see Chapter 6), then no AC function is activated. Function The DC voltage output may then be programmed from 0 AC with DC Offset When DC offset is added to any AC function, there are minimum and maximum offset limits which must be observed.
Page 70
Table 5-4. Maximum DC Offset with any AC Function Attenuation Entry DC Offset DC Of&et Factor peak-to-peak) Entry (+ or -) 1.000 mV w i t h 4.500 A = 1000 0.001 mV 3.333 mV with 3.333 mV 3.334 mV with 14.99 mV A = 300...
When the high-voltage output is used (option 002 is Option installed), the minimum and maximum permissible DC offset voltages may be determined by multiplying the amplitude and offset values in Table 5-4 by four. The equation given on the previous page must be changed to: Maximum DC offset = (20/A) - (Amptd/‘L) where A = Attenuation factor (from Table 5-4) and Amptd = Amplit,ude...
“EXT-REF” is shown in the display if an external reference frequency is used. Changing the phase of the HP 3324A will then cause the phase between the two instruments to change accordingly. Note In this case the phases of the two instruments are not calibrated, just locked.
To change the phase, either enter a new value with the numeric keypad, and press the unit key (den. To enter a negative phase, press modify the existing value using the modify function. edit the existing value using the edit function. Note For squarewave frequencies below 25 KHz, phase changes greater than 25”...
Calibration respect to the other. To be able to do this it is necessary to have one of the HP 3324As installed with option 003 (slave), and the other with option 004 (master). If one of the instruments has option 001, High-Stability Frequency...
111’ 332412’s containing options 003 and 004. If the two cables a.re the same then in both HP 3324A’s the “Sig.d/ps:” zero. To account for a difference in the cables enter the delay in one instrument with respect to the other, that is enter the delay in the one with the longer cable.
Page 76
DUT with an oscilloscope. replace Keystrokes Result Press this on both HP 3324A’s, so that the utilities menu is accessed. Scroll through the menu on both HP 3324A’s until “set default” is shown. Scroll through the menu on both HP 3324A’s until “Sig.d/ps: Press this only on the slave instrument, so that an automatic phase calibration is performed.
Changing the Calibration Cable Delay (003 or When an automatic phase calibration option 004) is installed “Cal.d/ps: n”, where n is a value from 0 to 59999, will be displayed in the UTILS menu. This can be used to account for the delay caused by changing the standard cable (supplied) for connecting the Phase Cal Out of the master (option 004) to the Phase Cal In of the slave (option 003).
Waveforms Introduction This chapter explains how to use the (E) key, which is one of the Menu keys. Press (jj to display the available waveforms, as shown in Figure 6-l. The meaning of each of the symbols can be seen in Figure 6-2. The waveforms always shown are the sinewave, squarewave, triangle, positive slope ramp, negative slope ramp, auxiliary TTL...
Page 80
T T L O u t p u t flux Figure 6-2. Waveforms Available Once a waveform has been activated, the HP 3324A jumps back to the point (menu) previous to entering the waveform menu, except for the following cases: the “AUX”...
The auxiliary TTL clock waveform, chosen by “RUX”, is an output that is compatible to TTL levels. Into a into an open impedance, the high voltage-level is V. When “AUX” is selected, the main signal output is switched off. A TTL clock signal is output from the sync connectors for frequencies from 1 mHz to 60 MHz.
Sweep Introduction Pressing the (G) key, one of the Menus keys, displays the Sweep menu, an example of which is shown in Figure 7-1. Mode I N T ERVRL CON’ M R R K E R STOP-FREQ SWEEP-TIME U s e d t o ed\t t h e P a r a m e t e r s t h e I n t e r v a l M a r k e r...
Page 84
This is also the case at power-on, if the HP 3324A is in the HP 3325A compatible mode. If not, the instrument is powered up to the settings that were at power-down. 7-2 Sweep...
Page 85
Table 7-1. Sweep Default State Parameter Mode Marker/Interval 10 (0 t,o 9) No. of stores Retrace Time Multi-Interval Sweep: 1 (liilear, all int,ervnls) Sweep Mode 1 M!lz (all intervals) Start Frequency Stop Frequency 5 Mlllz (all intervals) Marker Frequency 1 xc. (all inhvals) Sweep Time Activated Multi-Marker Sweep:...
Note Changing the number of intervals will cause the HP 3324A to reset all sweep intervals to their default state, and all stores to be cleared, so don’t program any intervals until you are sure you have enough available.
Page 87
Each interval is programmed with its own start frequency, stop frequency, sweep time, marker frequency, and whether its swept either linearly or logarithmically. means A linear interval may be either up or down, this that the stop frequency may be higher or lower than the start frequency.
Page 88
The marker output is a TTL-compatible output. It is high at the start of each interval, and when the marker frequency is reached it goes low, where it remains until the end of the interval. 50 individually-programmmed intervals may be used in a customized sweep-sequence of upto 100 intervals.
Page 89
The sweep sequence (which can consist of one interval) can be swept either singularly or continuously. A single sweep means that the HP 3324A sweeps from the start frequency of the first interval in the programmed sequence to the stop frequency of the last interval in the sequence, and then remains there.
This section will show how to set up a multi-interval Setting up a sweep sequence. Press lsweepl to show the Sweep menu. Multi-Interval To choose the multi-interval mode, put the cursor on Sequence “INTERVAL” and press (Select). A display similar to that in Figure 7-4 will be shown.
Page 91
When “INTERVAL” has been selected the current sweep sequence is shown. In the above example that is 1~3+6~2+9~4~9. This means that the sequence 1 3 4 5 6 2 3 4 5 6 7 8 9 4 9 is swept. m means “and then”, while (3) means “through”.
Page 92
To do this press the a key. You will then move onto some point in the list of start frequency, stop frequency, and so on. The position depends on where the cursor was when you last left and is shown by the parameter blinking.
Example This example shows how to enter the sweep menu, and set up a continuous multi-interval sweep containing three sweep intervals. The intervals will be swept in the order 2, 3, 1, and the retrace time will be 5 s. The first thing to do is to enter the Sweep menu, choose the multi-interval mode, and set up the first interval, which will have a start frequency of 1 KHz, a stop...
Page 94
The following figure shows what the display will look like now : I N T E R V R L S I N G L E C O N T M A R K E R - F R E Q SWEEP-TIM Figure 7-5.
Page 95
The following figure shows what the display will look like now: STOP-FREQ MRRKER-FREO S W E E P - T I M E Figure 7-6. Example of 2nd Interval The third interval must now be programmed. It will have a start frequency of 5 KHz, a stop frequency of 1 KHz, a marker frequency of 3 KHz, a sweep time of 2 s and will be swept linearly.
Page 96
The following figure shows what the display will look like now: S I N G L E C O N T MARKER-FREQ SWEEP-TIM Figure 7-7. Example of 3rd Interval Now the sequence must be set, the continuous mode chosen with a retrace time of 5 s, and the sweep started. Keystrokes Enter The current retrace time is shown, and a continuous mode...
Multi-Marker The multi-marker sweep consists of only one linear interval, but upto 9 markers. The frequency for each Sweep marker is programmed separately. You can then activate the markers that you require. The marker pulsewidth is approximately 1 ms. For example consider having programmed the following interval.
Page 98
Figure 7-9. An Example Marker Sequence There is a minimal time difference of 1.5 ms between start- and marker-, marker- and marker-, and marker- and stop-frequencies. If a marker is too close to the start- or stop-frequency, it is not generated. If two markers are too close the second one (in time) is not generated.
A continuous sweep means that the HP 3324A sweeps from the start frequency of the interval to the stop frequency, and then remains there for a programmed time (retrace time), before starting again at the start frequency.
Page 100
In the sweep display, two values in the form “n of 9” are shown (in the above example “n” = “1”). “n” means the number of the marker whose frequency you will see if you put the cursor onto the marker frequency. The “9” is the number of markers that can be programmed and activated in the multi-marker interval.
Page 101
Having chosen the marker number you wish to program, you can move onto changing the frequency. To do this press (5J. You will then move onto the marker frequency. You can then change the frequency by either: Typing in a new value with the numeric keypad, and then pressing the appropriate unit key.
Example This example shows how to enter the sweep menu, and set up a single multi-marker sweep containing three markers. The start frequency will be 1 KHz, the stop frequency 10 KHz, and the sweep time 10 s. The markers will have the frequencies 3 KHz, 5 KHz and 7 KHz.
Page 103
The following display will be shown: SINGLE C O N T NRRKER-FREP SWEEP-TIME Figure 7-11. Marker Interval Example A single sweep has to be chosen now and the sweep started. Result Keystrokes T h e s w e e p s t a r t e d .
This section explains the use of the sweep keys placed to The Sweep Keys the right of the front panel. Figure 7-12. The Sweep Keys Once a sweep has been defined it can then be started. This is done using the (ResetStartJ key. The (Start) key performs three functions for the sweep operations: (Reset) cancels the sweep.
If a sweep has been stopped, reset or started can be seen on the right of the display, where an arrow is shown next to either “Reset”, “Run” or “Stop”. In the multi-interval linear sweep mode, the (afx2) bandwidth, respectively. If one of the absolute sweep frequency limits (0 and ‘21 MHz for a sinewave) is exceeded, the new interval is clipped.
The instrument’s parameters are set to their default values if “set default” is displayed and m is ‘pressed. Parameters The default parameters are shown in Table 8-1. Note Setting the HP 3324A to default will cause the instrument to clear all its stores. Utilities 8-1...
Page 108
Table 8-1. Parameter Default Values Default Parameter Function Sine Frequency 1000 Hz Amplitude Phase 0 degrees DC Offset Selected Output 1 (front) output Marker/Interval Mode 1 (multi-interval) No. of stores 10 (0 to 9) Retrace Time 0 sec. (auto-retrace) Multi-Interval Sweep: Sweep Mode 1 (linear, all intervals) Start Frequency...
40 - 50 Note Changing either the number of stores or intervals will cause the HP 3324A to reset all sweep intervals to their default state, and all stores to be cleared. Recommendation: Use this command sparingly and only when absolutely necessary.
HP-IB address. The default HP-111 address is 17. Choosing HP The HP 3324A can be set so that at power-on, it powers up just like the HP 3325A. This means that in the 3325A compatible mode the main output signal is enabled and...
The “all” test is also an interactive test. If an error is shown return the instrument to an HP Service Center for repair. To perform one of the self-tests, display “selft: std./all”...
Checking the In the utilities menu, one of the parts of the menu shows the software revision incorporated in the HP 3324A. Software Revision Scroll the menu until “ROM rev. : x.x” is displayed. This is the software revision. Switching the If the high-voltage option 002 is installed, it can be either enabled or disabled as the main output signal.
A complete syntax list of the HP 3324A programming commands can be found in Appendix E. The HP 3324A The HP 3324A interfaces to the HP-IB as defined by the IEEE Standard 488.1. The interface functional subset HP-IB Capabilities which the HP 3324A implements is specified in Table 9-l.
HP 3325A. Note Execution times for the HP 3324A HP-IB commands are slower than those for the HP 3325A. To make sure that a command has been processed, check the “Busy” bit in the status byte. Chapter 11 and Appendix E show which HP 3324A commands are compatible to the HP 3325A and which are not.
Page 117
Frequency 10 MHz Stop Frequency 5 MHz (all markers) Marker Frequency 1 sec. Sweep Time I (only the first marker) Activated Markers (unbuffered) Data Transfer Mode SRQ Mask FR (see section “Omitting Default Mnemonic Mnemonic”) Programming the HP 3324A 9-3...
HP-IB Display The following list explains the indicators that can be seen in the display when the HP 3324A is being Indicators controlled remotely. “REMOTE” “REMOTE” is displayed when the HP 3324A is operating under remote control. While in remote (and local lockout is not in effect), only the @ key is recognized.
Table 9-3. Accepted Characters Description Character White spaces HT, LF, CR, space Block terminator Plus sign Minus sign / Range separator Item separator Decimal point Digits / Integer List terminator Letters Programming the HP 3324A 9-5...
HP-IB Messages Table 9-4 shows which HP-IB messages are implemented/not implemented in the interpreter. Table 9-4. HP-IB Messages Treatment Command Data byte Implemented as usual Device clear Sets 3324 into reset state, see section “The Interpreter” End or identify End is treated as white space...
Command Groups The HP 33‘24A works in two separate data transfer modes. In mode 1, the unbuffered mode, commands and Data Transfer are executed as soon as they are received. In mode Modes 2, the buffered mode, the execution control depends on the command group to which a command belongs.
Page 122
Treatment Group Executed immediately, HP 3324A is put into mode 2. Group 0 Group 1 Command is remembered but not executed. Group 1 Command is remembered but not executed. Group 0 Commands “FUl” and “FR12.3MH” are executed, along with command “PH5DF.
Page 123
In data transfer mode 1, the only command that will not be executed is the string The following events will cause remembered, but not executed commands to be lost: power off remote commands. Programming the HP 3324A 9-9...
1, and 3 MHz. This could be done in the following way: Send program string Do measurement Send program string “2MH” Do measurement Send program string “3MH” Do measurement 9-10 Programming the HP 3324A...
To change the address, type in the required address with the keypad, and then press IEnter). The HP-IB refuses to accept a new HP-IB address if it is either in the remote state, or if it is addressed either as a listener or as a talker.
Status Reporting Introduction Another important feature of the III’ 3324A is that you can program it to interrupt the controller when certain status/error conditions are met. An 8-bit register, the Status Byte, contains the instrument’s status. In order for an interrupt to occur, it must be programmed via the Service Request Enable Mask.
Page 128
7 Bit 5 Bit 4 Bit. 3 Bnt 2 Bit L L-- I =Prog P am s:r',b"p I =Sweep Stopped S t a r t e d I-System Always 0 I ” P r o g r e s s Reqest Busy Figure 10-l.
Page 129
Sweep. Set when a sweep is in progress, clear when a sweep is in progress. Cannot be configured to cause an SRQ. Require Service. Set when the HP 3324A requires service (sent SRQ). 1 2 8 Busy. Set while a command is being executed, clear when instrument is not busy.
Page 130
The program string error bit is set when a command string is received that the interpreter does not understand, or when a command is not executable for When the “IER” query some reason. is sent, an error number will be returned, with the following meanings: Error Error Description...
Page 131
The system failure bit is set when a system failure is encountered. The specific error may be found by sending the “ISE” query, and evaluating the returned error number: System Failure Description System Failure Number Amplitude calibration failed Phase calibration failed External reference unlocked Main oscillator unlocked Self-test failed (report actual number to...
Forcing an SRQ An SRQ is forced when one of the above bits in the Status Byte Register goes from “0-> 1” AND the corresponding Service Request Enable Mask bit has been set. Bits O-3 are never set if they are not enabled via the SRQ-mask command.
Service Request The SRQ mask can be set to mask bits O-3 of the Status Register. To set the mask first determine which Enable conditions you want to interrupt the controller, for example sweep started, sweep stopped and so on. Determine the decimal number (1-16) corresponding to those conditions.
Example 1 The following example shows how to set the SRQ mask, so that if a “0 + 1” transition of the Sweep Started or System Failure bit occurs, an interrupt is sent. S y s t e m Sweep A l w a y s S y s t e m Sweep Sweep...
Example 2 order to see when a sweep has started. Result Program Clears all devices. HP-IB SEND 7; CMD 20 Sets SRQ mask for bits 2 and 3 of the status OUTPUT 717; "MSL" byte. Serial polls the status byte.
Program end. The first thing to check if the amplitude is not shown, is that the HP-IB address set on the HP 3324A is the same as that you are using (in this case 17). HP-IB Commands and Programming Examples...
Page 138
Table 11-l. HP 3324A Command Summary S R O - 9 Recall set SNR Integer White C M O - o f f node 1 - on Function F U O - D C 1 - Sine 2 - Square...
Page 139
Table 1 l-l. HP 3324A Command Summary (continued) 0 phase Select R F 0 - f r o n t o u t p u t 1 - rear o u t p u t O O F O - o f f...
Page 140
Table 11-l. HP 3324A Command Summary (continued) Data 3325 Multi-Interva start freq Number See FR Index, number stop freq Number Index, number marker freq Number Index, number sweep time Number SE - sec. Index, number sweep mode 1 - lin...
N o t e s (1) The meaning of the command has been changed. See the section “Command Groups and Data Transfer Modes” in Chapter 9. The data transfer command, MD, allows you to select between a buffered and an unbuffered transfer mode. In the unbuffered mode, the command is executed immediately.
Page 142
33.33 mV with 33.33 mV 149.9 mV 33.34 mV with 0.010 mV 99.99 mV with 116.6 mV with 450.0 mV 100.0 mV 0.100 mV 333.3 mV with 333.3 mV 333.4 mV with 1.499 V 11-6 HP-IB Commands and Programming Examples...
Page 143
9. The list must define the subset of enabled markers (not the sequence). All list elements must be in the range to 9. The default list is “1” (only one marker, the first one). HP-IB Commands and Programming Examples 11-7...
Page 144
In all other cases, the retrace time is evaluated according to (13). The output frequency sweeps down/up from the stop frequency of the last interval in the sweep sequence to the start frequency of the first interval. 11-8 HP-IB Commands and Programming Examples...
Page 145
(12) If the total sweep time (for multi-intervals, this is the sum over all selected intervals) is greater than 99.99 seconds, then the X-drive is switched off. This means that the X-drive output is held low all of the time. The Z-blank output, however, works as usual.
All of the programs assume the following: an HP 9000, Series 200 or 300 Computer as controller 2. that BASIC is the programming language 3. that the HP 3324A is preset to HP-IB address 17 Function Generator The following program shows an example of how to program the HP 3324A as a function generator.
Multi-Interval Sweep The following program shows an example of how to program the HP 3324A as a multi-interval sweep Generator generator. The program assumes that the amplitude, waveform, data transfer mode, DC offset, phase, output and no. of intervals are the same as set in the previous example.
Multi-Marker Sweep The following program shows an example of how to program the HP 3324A as a multi-marker sweep Generator generator. The program assumes that the previous two example programs have been run directly beforehand. The program sets the following information.
Introduction The procedures in this chapter test the performance of the instrument. The complete specifications to which the HP 3324A is tested are given in Appendix A. All tests can be performed without access to the interior of the instrument.
If the HP 3324A fails any performance test, return the Test Failure instrument to the nearest Hewlett-Packard Sales/Service Office for repair. Specifications are the performance characteristics of the Instruments instrument which are certified. These specifications, Specification listed in Appendix A, are the performance standards or limits against which the HP 3324A can be tested.
Page 153
Table 12-1. Recommended Test Equipment Recommended Model Instrument Critical Specifications HP 1725A Vertical A n a l o g Oscilloscope Bandwidth: DC to 275 MHz Deflection: 0.01 to 5 V/div Horizontal Sweep: 10 ns to 0.5 s/div Delayed Sweep HP 54111D...
Page 154
Ranges: 0.1 to 1 V HP 400FL/3400A AC Voltmeter Frequency Range: 20 Hz - 1 MHz Input Impedance: 11 MR Meter: Log scale Frequency: 20 KHz HP 3324A/3325B Sincwave Signal Source Amplitude: 10 V(p-p) into 500 HP 6214A/6214B DC Power Supply Amps: 10 mA...
Verify that all harmonics are at least 25 dB below the fundamental. 4. Set the HP 3324A to 15 MHz and verify that all harmonics are at least 25 dB below the fundamental. Performance Tests 12-5...
1.99 MHz 200 KHz - 4 0 dB 6. Set the HP 3324A frequency to 50 KHz and the amplitude to 9.99 mV(p-p). 7. Set the spectrum analyzer controls to display the fundamental and at least three harmonics. (It may be necessary to decrease the video bandwidth to separate the harmonics from the noise floor.) Verify that all...
Page 157
4 7 5 O h m s Figure 12-1. Harmonic Distortion Verification Test Set-Up (High-Voltage Output) 2. Select the high-voltage output on the HP 3324A. Set the amplitude to 40 V(p-p) and the frequency to 100 3. Set the spectrum analyzer controls to display the fundamental and at least three harmonics.
Equipment Required Spectrum Analyzer Mixer Spurious 1. Connect the HP 3324A signal output to the spectrum analyzer 500 (R,F) input and the III’ 3324A EX’I Procedure REF input to the analyzer 10 MHz reference output, as shown in Figure 12-2.
Page 159
25.998 MHz to display the 3:2 mixer spur. Verify that this spur is at least 55 dB below the fundamental. 7. In a similar manner, change the HP 3324A frequency and the spectrum analyzer center frequency to the following frequencies. For each setting, verify that all spurious signals are 55 dB below the fundamental.
Close-in Spurious This procedure continues on from the previous one. (Fractional N Spurs) 1. Set the HP 3324A to 5.001 MHz and the amplitude to Procedure 448.3 mV(p-p). 2. Set the spectrum analyzer controls as follows: 5.001 MHz Center Frequency...
1 MHz Low Pass Filter 1. Connect the equipment as shown in Figure 12-3, Procedure connecting the 15 KHz noise equivalent filter output to the AC voltmeter. Phase lock the HP 3324A and the signal generator together. HP 3324A Voltmeter...
Page 162
5. Change the HP 3324A frequency to 19.9 MHz. 6. Connect the 15 KHz filter output to the digital voltmeter. 7. Press the HP 3324A (Phase) key. Using the modify keys, adjust the output phase for minimum reading on the digital voltmeter.
Ramp Retrace This procedure tests the HP 3324A retrace time of the positive and negative slope ramps. Time Specifications Equipment Required Digitizing Oscilloscope Procedure 1. Connect the HP 3324A signal output to the oscilloscope vertical input. (If your oscilloscope does not have a 5052 input, use a 5052 feedthru termination at the input.)
Equipment Required Electronic counter (calibrated within three months or with an accurate 10 MHz external reference input) 1. Connect the HP 3324A signal output to the electronic Procedure counter channel A input with a 500 fecdthru termination. Allow the HP 3324A t,o warm up for 20 minutes and the counter’s frequency reference to...
Set the counter to average 1000 periods. The electronic counter should indicate 6. Change the HP 3324A function to a positive slope ramp. The electronic counter should indicate This procedure compares the phase increment accuracy Phase Increment to the specification.
Page 166
Slope A and 1~ Positive Sample Rate Maximum 5. Press the III’ 3324A m key to display phase. Using the modify keys, adjust the pha.se until the counter reads approximately 200 ns. Press the blue (Shift key, and then the key.
Page 167
8 and 6, and record in the Time Difference column. The difference should be f0.5”. 10. Set the HP 3324A phase to -10”. 11. Reset the counter. Record the counter reading in the space for 10’ Increment Time Interval.
Note For each new setting of the amplitude, press @&i-G” to perform a.n amplitude calibration. 1. Sinewave Test. Connect the HP 3324A signal output Amplitude Accuracy through a 50R feedthru termination to the AC at Frequencies upto digital voltmeter input.
Page 169
5. Change the HP 3324A amplitude to 1.061 Vrms (3 KHz and 100 KHz as above. Verify that all three voltmeter readings are between 1.037 and 1.085 Vrms (ho.2 dB). 6. Change the HP 3324A amplitude to 0.3536 Vrms (1 V(p-p)) and set the DC offset to 1 mV. Set the HP 3324A frequency to 100 Hz, 1 KHz and 100 KHz, and read the AC voltage.
Page 170
Add the two readings to obtain volts peak-to-peak. Verify that the sum is between 9.85 and 10.15 V. 14. Change the HP 3324A function to a triangle. 15. Adjust the oscilloscope delay to read the positive peak voltage on the high-speed digital voltmeter.
Page 171
Verify that the sum of the absolute values is between 9.85 and 10.15 V. 22. Change the HP 3324A function to a triangle and the frequency to 2 KHz. Set the oscilloscope main sweep to 50 PsJdiv. Adjust the delay and position. Set the positive and negative trigger to read the peaks.
Page 172
23. Change the HP 3324A function to a positive ramp and the frequency to 500 Hz. Set the oscilloscope main sweep to 0.2 ms/div (adjust the sweep vernier to return the peaks to the center screen). Verify that the voltage is between 9.85 and 10.15 V.
Page 173
29. Change the HP 3324A amplitude to 3 V(p-p), and press [mj. Set the HP 3324A frequency to 99.9 Hz and the function to squarewave. 30. Repeat tests 9 through 28. Test limits are as follows: Function 99.9 Hz Square 3.045 v...
Page 174
31. Change the HP 3324A amplitude to 1 V(p-p), press frequency to 99.9 Hz and the function to square. Set the oscilloscope’s vertical sensitivity to 0.05 V/div for all 1 V(p-p) tests. 32. Repeat tests 9 through 28. Test limits are as follows:...
AC voltmeter with a 6-foot cable. Connect a line. Select the high-voltage output on the HP 3324A. “HIGH-VOLT” is shown in the display. Set the HP 3324A function to a sinewave, the 14.14 frequency to 2 KHz, and the amplitude to Vrms reading should be between 13.86 and 14.42 Vrms.
Page 176
Connect the tap to the input of the high-speed voltmeter, as shown in Figure 12-6. 6. Set the HP 3324A frequency to 2 KHz and the amplitude to 40 V(p-p). Set the voltmeter to the 1 V range and external trigger. Set the oscilloscope as...
Frequency DC Offset 10 dBm Amplitude 2. Connect the HP 3324A signal output via a 500 to measure AC voltage. 3. Press [Ampl followed by C-Cal). Wait 2 seconds and record the voltmeter reading as Vnrms. 4. Connect the equipment as shown in Figure 12-7 (the...
Page 178
Calculate the amplitude accuracy using the following equation: 8. Verify that all results are within 9.4 dBm and 10.6 9. Change the HP 3324A frequency to 1 KHz and the amplitude to 18 dBm. 10. Repeat steps 2 to 7 and verify that all results are within 17.6 and 18.4 dBm (accuracy = 60.4 dB).
Sensitivity 10 V/div Time/div 1 ms 3. Set the HP 3324A to 40 V(p-p) sinewave and 1 KHz. Adjust the oscilloscope intensity and focus for a sharp trace. 4. Use the modify keys to increase the HP 3324A frequency from 1 KHz to 1.001 MHz in 200 KHz steps.
Connect the HP 3324A signal output directly to the the DC digital voltmeter input. 3. Set the HP 3324A DC offset to 5 V, then press 4. The voltmeter reading should be between i-4.950 and 5. Change the HP 3324A DC offset to -5 V. The voltmeter reading should be between -4.950 and...
HP 3324A output directly to the voltmeter input. 2. Select the high-voltage output on the HP 3324A. “HIGH-VOLT” will be shown in the display. 3. Set the HP 3324A DC offset to 20 V. The voltmeter reading should be +19.775 to 20.225 V. Performance Tests 12-31...
DC Offset This procedure tests the HP 3324A DC offset accuracy when an AC function output is present. Accuracy with AC Functions Specifications DC + AC, upto 1 MHz: 1.2% For ramps upto 10 KHz: 2.4% DC + AC, from 1 MHz to 20 MHz: 3%...
Page 183
5. .Change the HP 3324A frequency to 999.9 KHz. The voltmeter reading should be -4.440 to -4.560 Vdc. 6. Change the HP 3324A DC offset to +4.5 V. The voltmeter reading should be +4.440 to +4.560 Vdc. 7. Set the HP 3324A function to square. The voltmeter reading should be -f-4.440 to i-4.560 Vdc.
Triangle Linearity This procedure tests the linearity of the HP 3324A triangle wave output. As the triangle and ramp outputs are generated by the same circuits, this procedure effectively tests the ramp linearity also. Specifications Equipment Required High-speed DC Digital Voltmeter...
Page 185
Number of readings Trigger External Note The HP 3437A triggers on the negative going edge of the HP 3324A sync squarewave. 4. Set the voltmeter delay to 0.00003 (seconds). Record the voltmeter reading on the Performance Test Positive Slope Measurement, (lO%)yl.
Page 186
6. Measure the voltage at each 10% segment point on the negative slope by setting the voltmeter delay to the following. Enter on the Performance Test Record in the appropriate spaces under Negative Slope Measurement. Percent of Slope Delay 0.00008 0.000085 0.00009 0.000095...
Page 187
11. First determine the value of al using the following equation: where Cx, Cy, Cxy, CxCy, ,Xx2, and (CX)~, are the previously calculated values entered on the Performance Test Record. where n=9 (the number of points to be calculated). 12. Determine the value of ac using the equation: 13.
The voltages measured and recorded in the within the calculated tolerances. Ramp Period This procedure tests the variation between alternate cycles of the HP 3324A positive and negative slope Variation ramps. Specifications < kl% of period, maximum...
Page 189
6. The reset portion should show more than one line, as in Figure 12-9. The lines should not be separated by more than five divisions on the display. 7. Change the HP 332412 function to positive slope ramp and set the oscilloscope trigger to negative to verify the positive ramp.
> 1 MHz - 10 MHz ~t1.2~ , tYP. Digitizing Oscilloscope Equipment Required Note Connect a 332442 signal to the oscilloscope and using 15109, an adder (splitter) HP check the oscilloscope interchannel accuracy and take the error into account. 12-40 Performance Tests...
CRL IN 0 0 0 3 SLRVE BNC CRBLE 500HM S A M E L E N G T H Figure 12-10. Phase Calibration Test Set-Up 2. Set the two HP 3324As as follows: Util menu Set default Frequency 1 MHz...
Page 192
# Avgs Press (Edge). Check that & < 4.17 ns. Change the waveform of the two HP 3324As to square. Select the Delta V menu on the scope and with preset levels at 50-50% press Auto Level Set.
Test Record 1 Hewlett-Packard Model 332411 Teat Performed By Synthesized Function/Sweep Generator Date Serial Number Comments CSO No. Pass 0 Fail 0 Harmonic Distortion Test Specification Fundamental F’requency -25 dB 20 MHz 15 ME&z 14.9 MHz -30 dB 2 MHz 1.99 MHz 200 KHz 10 KHz...
Test Record 2 Hewlett-Packard Model Test Performed By 3324A Date Synthesized Function/Sweep Generator Comments Serial Number CSO No. Customer Spurious Signal Test Pass 0 Fail 0 Spurious Mixer 2.001 4.100 MHz 6.100 MHz 8.100 MHz 10.100 MHz 12.100 MHz 14.100 MHz 16.100 MHz...
Test Record 4 Hewlett-Packard Model 3324A Test Performed By Synthesized Function/Sweep Generator Date Serial Number Comments Customer CSO No. Frequency Accuracy Test Pass 0 Fail 0 Sine, 20 MHz Square, 10 MHz - - - - - - - - - - - - - - - - - - - - - - - - -...
Page 197
Test Record 5 Hewlett-Packard Model 3324A Test Performed Hy Synthesized Function/Sweep Generator Date Serial Number Comments Customer CSO No. Amplitude Accuracy Test Pass 0 Fail 0 Measured Amplitude: 3.536 Vrms Sine, 100 Hz: 3.455 v 3.617 V Sine, 1 KHz: 3.455 v...
Page 198
(2 of 4) Test Record 5 Amplitude Accuracy Test (continued) Function Test Measured Amplitude: 10 V(p-p) Square, 99.9 Hz 10.15 v 9.95 v Triangle, 99.9 Hz 10.15 v 9.85 v Pos. Ramp, 99.9 Hz 10.15 v 9.85 v Neg. Ramp, 99.9 Hz 10.15 v 9.85 v 10.15 v...
Page 199
Test Record 5 Amplitude Accuracy Test (continued) Function Test Measured Amplitude: 1 V(p-p) DC: 1 mV Square, 99.9 Hz 1.022 v 0.978 V 1.027 V Triangle, 99.9 Hz 0.973 v 1.027 V Pos. Ramp, 99.9 Hz 0.973 v 1.027 V Neg.
Page 200
Test Record 5 Accuracy Amplitude Test (continued) Min. Amplitude Flatness Test 10.2 dBm 9.8 dBm Sine, 10 dBm, 1 KHz 10.6 dBm 9.4 dBm 101 KHz 10.6 dBm 9.4 dBm 500 KHz 10.6 dBm 9.4 dBm 1 MHz 10.6 dBm 5 MHz 9.4 dBm 10.6 dBm...
Page 201
Test Record 6 Test Performed By Hewlett-Packard Model 3324A Function/Sweep Generator Date Comments Serial Number Customer CSO No. Pass 0 Fail 0 DC Offset Accuracy Test (DC Only) Minimum Maximum Measured -5 v -4.950 v - - - - - - - - - - - - - - - - -5.050 v 1.499 v...
Page 202
Test Record 7 Test Performed By Hewlett-Packard Model 3324A Synthesized Function/Sweep Generator Date Serial Number Comments Customer CSO No. Offset Accuracy Test with AC Functions Pass 0 Fail 0 Measured Maximum Minimum Sine, 21 MHz 4.5 v -4.5 v -4.350 v -4.650...
Page 203
Test Record 8 Test Performed By Hewlett-Packard Model 3324A Date Synthesized Function/Sweep Generator Comments Serial Number CSO No. Customer Pass 0 Fail 0 Triangle Linearity Test Minimum Maximum Calculated x Values Positive Slope x times y Tolerance Tolerance Best Fit...
Page 204
Test Record 8 Pass 0 Fail 0 Triangle Linearity Test (continued) Minimum Maximum Calculated Negative Slope Tolerance Tolerance Best Fit Measurement Straight Line (10%) y1----- - -- - - - - - - (20%) y2----- x3 = 3 (30%) y3----- (40%) y4----- x5 -= 5 (50%) y5-----...
Page 205
Test Record 9 3324A Test Performed 13~ Synthesized Function/Sweep Generator Date Comments CSO No. Automatic Phase Calibration Test Pass 0 Fail 0 Formula: At < 4.17 Sine, 1 MHz At < 2.78 ns 1 MI12 Performance Tests 12-55...
0°C to 55°C ambient temperature. All specifications describe the warranted performance, except for the Main Signal Output, Squarewave Characteristics (also by Option 002), Auxiliary Outputs, Auxiliary Inputs, HP-IB Control and General specifications, which describe the typical performance. Waveforms Sine, square, triangle, negative and positive ramps, DC and TTL clock.
Accuracy: calibration with standard frequency reference. Stability: option 001, high-stability frequency reference). Warm-up Time: 20 minutes to within specified accuracy. Impedance: Main Signal output Return Loss: (Supplementary) > 3 V, 5 MHz to 20 MHz. Floating: Output may be floated upto 42 V peak (AC + DC). Connector: BNC, switchable between front and rear panel.
Page 209
1 mV - 2.999 mV 100 mV - 299.9 mV 3 mV - 9.999 mV 300 mV - 999.9 mV 10 mV - 29.99 mV 1 v - 2.999 v 30 mV - 99.99 mV 3 v - 10.00 v Resolution: 4 digits (0.03% of full range).
Squarewave: 0.001 Hz 100 KHz >lOO KHz 10 MHz Triangle: 0.001 Hz to 2 KHz >2 KHz to 10 KHz Ramps: 0.001 Hz to 500 Hz >500 Hz to 10 KHz With DC offset, increase all sinewave tolerances by 0.2 Phase Noise: -50 dBc for a 30 KHz band centered on a 20 MHz Purity...
Harmonically related signals will be less than the following levels relative to the fundamental: Frequency Range Harmonic Level 0.1 Hz - 199 KHz -60 dBc 200 KHz - 1.99 MHz -40 dBc 2 MHz - 14.9 MHz -30 dBc 15 MHz - 20 MHz -25 dBc Rise/Fall Time:...
DC Offset DC only (no AC signal): 0 to f5 V / 50Q DC + AC: Maximum DC offset f4.5 V on highest range; decreasing to f4.5 mV on lowest range. Resolution: digits Accuracy: DC only: f0.015 mV to f50 mV, depends on offset chosen, ~4~0.02 mV.
Sequence Modes: Frequency Sweep Sweep Single, continuous. Sweep Function Modes: Multi-Interval: 50 different intervals can be sequenced and repeated in any order in a sequence which can contain Frequency-switching-time between intervals (to within 1 Hz): Multi-Marker: sweep (no Multi-Interval) is performed. Reaching a marker frequency generates a high-to-low transition at the marker output.
Auxiliary Outputs SYNC Output: 1 mHz to 21 MHz phase synchronous squarewave with (Supplementary) the same frequency as the main signal output, or 1 mHz to 60 MHz Auxiliary TTL Clock (main signal output switched off). Output impedance: 5051 Output levels: high level > 1.2 V, low level < 0.2 V Connector: BNC, front and rear panels.
Output amplitude: 0 dBm Connector: rear panel BNC. Reference Input Auxiliary Inputs For phase locking the HP 3324A to an external (Supplementary) frequency reference. Signal from 0 dBm to 20 dBm into 500 Reference signal must be a sub-harmonic of 10 MHz from 1 MHz to 10 MHz.
25OC, for an off time of less than 24 hours. 10 MHz Oven Output: 10 MHz squarewave for phase-locking additional instruments to the HP 3324A. Output impedance: 50R Output level: >4.5 dBm Connector: rear panel BNC.
HP 3324As without Options external measuring instruments. Option 003 has to be installed into one HP 3324A, option 004 has to be installed into the other instrument. Adjust factors for different cable delays between master/slave and can be entered in the UTILit,y menu.
Page 218
Phase Error The phase error is measured between the main signal outputs on the rear panel with either unequal or equal amplitudes and without DC offset. Phase is defined as the difference in rising edge to rising edge (measured with AC-coupled zero-crossing-points as reference points) for sine and squarewaves.
Operating Environment: General Temperature: 0°C to 55”C, if not otherwise stated. (Supplementar Relative humidity: 95%, 0°C to 40°C Storage temperature: -40°C to +75”C Power: V, &lo%; 48 to 66 Hz,; 100 VA max. Standby 20 VA max. Weight: 11 kg net, 16.5 kg shipping. Dimensions: 132.6 mm high x 425.5 mm wide x 497.5 mm deep (5.25 ins.
This appendix gives an explanation of the options and accessories available for the HP 3324A. Table B-l lists the options available for the HP 3324A. These options are available when the instrument is ordered by specifying the option number, or are available later by ordering the option part number (p/n).
001 requires 15 minutes of warm-up time to meet frequency specifications if power is disconnected for less then 24 hours. For more than 24 hours, the HP 3324A may require upto ‘72 hours warm-up time to meet the frequency specifications.
4 V(p-p). Turning the high-voltage off causes the output to be amplitude The HP 3324A specifications apply when the external load is 500R and the total capacitance is <500 pF. The same entry procedures and display features apply as for the standard configuration.
To be able to do this it is necessary Phase Calibration to have one of the HP 3324As installed with option 003 (slave), and the other with option 004 (master). If one of the instruments has option 001, High-Stability Frequency...
Page 225
l CWL IN Figure B-l. Automatic Phase Calibration Setup Pressing the (Phase] key on the slave (the one with option 003 installed) will cause a phase calibration between the two instruments to be performed. “phase is sl~own shortly in the display of the slave, before displaying what was previously shown.
Safety The Model HP 3324A is a Safety Class 1 instrument (instrument with an exposed metal chassis that is Considerations directly connected to earth via the power supply cable).
Power Requirements Caution Before applying AC line power to the HP 3324A, ensure that the voltage selector on the HP 3324A bottom panel is set for the proper line voltage and the correct line fuse is installed in the fuse holder. Procedures for changing the line voltage selector and fuse are contained in the following section “Line Voltage Selection”.
The line voltage selected for the HP 3324A is indicated by the switches on the bottom panel. Refer to Table C-l for the line voltage ranges and Table C-2 to set the line voltage and select the appropriate fuse.
Page 230
To change the line voltage and fuse: 1. Remove the power cord. 2. To check or replace the fuse, press the fuse holder in slightly, using a screw driver, and turn it to release the catch. Pull out the fuse holder. 3.
Power Cable In accordance with international safety standards, this instrument is equipped with a three-wire power cable. When connected to an appropriate AC power receptacle, this cable grounds the instrument cabinet. The type of power cable shipped with each instrument depends on the country of destination.
Page 232
Denmark Europe G r e a t Austral I a 8 1 2 0 - 1 8 120-2956 8 120- 2 0 - 1 3 5 1 8128-1378 S w i t z e r l a n d Figure C-2. Power Cables - Plug Identification The following work should be carried out by a qualified electrician - all local electrical codes being strictly observed.
HP part number 1251-0293. HP-IB Logic Levels The HP 3324A HP-IB lines use standard TTL logic, the levels being as follows: True = Low = digital ground or 0 Vdc to 0.4 Vdc, False = High = open or 2.5 Vdc to 5 Vdc.
S Y S T E M C O N T R O L L E R ) Figure C-3. HB-IB Connector The following summarizes HP 3324A operating Operating enviroment ranges. In order for the IIP 3324A to meet Environment specifications, the operating enviroment must be within these limits. Warning The HP 3324A is not designed for outdoor use.
BNC cable. For example, use HP p/n 5180-2459 (1.22 m, RG58V Triax, 500). Instrument Cooling The HP 3324A is equipped with a cooling fan mounted inside the rear panel. The instrument should be mounted so that air can freely circulate through it.
If the instrument is to be shipped to a Hewlett-Packard Return Shipments Sales/Service Office, attach a tag showing owner, return to HP address, model number and full serial number and the type of service required. The original shipping carton and packing material...
Introduction This appendix gives an overview of the Hewlett-Packard Interface Bus (HP-IB). Description of the The HP-IB is a bus structure that links the HP 3324A to desktop computers, minicomputers, and other HP-IB HP-IB controlled instruments to form automated measurement systems.
Page 238
Every participating device must be able to perform at least one of the following roles; talker, listener or controller. The HP 3324A can be either a talker or a listener. The full flexibility and power of the HP-IB is realized when a controller is added to the system.
Address Capability Primary addresses: 31 talk, 31 listen; secondary (2-byte) addresses: 961 talk, 961 listen. 1 talker and 14 listeners, maximum, at one time. The HP 3324A has only primary address capability. Table D-l lists the talk and listen HP-IB addresses.
This line is used by the active controller to define how information on the data lines should be interpreted by other devices on the bus. When ATN is low (true) the HP-IB is in the command and the data lines should be interpreted as bus mode...
The HP 3324A responds to a serial poll by putting its status byte on the data lines (see Chapter 10 Status and Event Reporting).
Page 242
Talk/Listen Each HP-IB device has at least one talk and one listen address unless the device is either totally transparent or Addresses is a talk-only or listen-only device. Device addresses are used by the active controller in the command mode (ATN true) to specify the talker (via a talk address) and the listener(s) ( via listen addresses).
The HP-IB interface system operates in one of two modes, controlled by the ATN bus management line: command mode (ATN true) or data mode (ATN false). (if an HP controller is used, the bus management lines are configured automatically and all necessary command strings are issued).
(front panel) control. If local lockout is not in effect, the HP 3324A responds by returning to front panel control. The Remote indicator on the front panel extinguishes if the HP 3324A is in Remote prior to the Local command. Local Lockout The local lockout command disables the m front panel key to avoid operator interference.
Bits 0, 1, 2, and 3 in the status byte may initiate an SRQ depending on the setting of the status byte mask. The status byte may be masked to select which of the four bits cause the HP 3324A to issue the SRQ. Trigger...
HP 3324A Command Syntax Introduction This appendix contains the syntax of the HP-IB commands for the HP 3324A. The list of commands in this appendix are given in alphabetical order. For more information see Chapter 9 Remote Control of the HP 3324A and Chapter 11 HP-IB Commands.
HP-IB Command The following table shows all of the HP-IB commands available, alphabetical order: List Table E-l. HP-IB Command List Description Description Mnemonic Phase Amplitude Calibration Recall State Amplitude Al’ Assign Zero Phase Rear or Front Signal Output H P 3325A Compatible...
Page 249
AC; Amplitude The AC command performs an amplitude calibration. If calibration is not successful, the FAIL bit of the status Calibration register is set. Command is available on the HP 3325A. Syntax HP 3324A Command Syntax E-3...
Page 250
The AM command sets the amplitude of the main signal. Sending IAM causes the instrument to output its current amplitude. Instrument Preset Value: 1.0 mV(p-p) Both commands are available on the HP 3325A. Syntax Table E-2. AM ‘cvalue” Range Given “units” “value” Range Description High Voltage 0.001 to 10.0...
Page 251
Zero AP; Assign The AP command assigns the current phase value to zero; subsequent changes in phase are referenced to this Phase point. Command is available on the HP 3325A. Syntax HP 3324A Command Syntax E-5...
Page 252
CM; HP 3325A The CM command switches the HP 3324A to the HP 3325A compatible mode or to the non-compatible mode Compatible at power-on. The difference between the two modes is explained in the following table: HP 3325A Compatible Mode...
Page 253
Table E-4. ICM Response Format Query Command Response Format HP 3324A Command Syntax E-7...
Page 254
IFR causes the instrument to output its current frequency. Instrument Preset value: 1 KHz Both commands are available on the HP 3325A. Syntax Table E-5. FR “value” Range Given “units” Table E-6. IFR Response Format...
Page 255
FU; Waveform The FU command selects the waveform function for the main signal output. Function Instrument Preset value: Both commands are available on the HP 3325A. Syntax Waveform Selections for “digit” Waveform DC only. Selects Selects Sinewave. Selects Squarewave. Selects Triangle.
Page 256
HV; High-Voltage The HV command controls the high-voltage amplifier option for the main signal output. output Instrument Preset value: 0 Command is available on the HP 3325A. Syntax Enable high-voltage amplifier. Table E-8. IHV Response Format Query Command Option Installed? Response Format...
Page 257
The query returns a one-digit Query code. If no error occurred, 0 is returned. Issuing the command clears the error code to 0. Command is available on the HP 3325A. Syntax Table E-9. IER Response Format Comman d Response Format...
Page 258
“0” or “00X” will be shown, where “0” means that the option is not installed and “00X” (X=option number) means that the option is installed. Command is not available on the HP 3325A. Syntax Table E-10. IOPT Response Format Response...
Page 259
The query returns a one-digit Query code. If no error occurred, 0 is returned. Issuing the command clears the error code to 0. Command is not available on the HP 3325A. Syntax Table E-11. ISE Response Format Query C ommand Response Format...
Page 260
RAM/ROM test failed (ROM=signature test) MFP (MC 68901) Test failed Device bus test failed Display test failed (display handshake) DAC test failed (analog voltage generation) FRAC-N chip test failed VCO test failed Sweep timer test failed Offset test, failed E-14 HP 3324A Command Syntax...
Page 261
HP 3324A program string has been accepted and all but the last character processed. In mode 2, the buffered mode, the execution control depends on the command group to which a command belongs.
Page 262
IXMF commands, respectively, when an index of 1 is used. To set and read the marker frequencies of other sweep intervals, use the XMF and IXMF commands. Instrument Preset value: 5.0 MHz Both commands are available on the HP 3325A. Syntax Table E-13. MF Range Given “units”...
Page 263
IMKS causes the instrument to output a list of the activated markers. Use the MMF command to set the frequency of the various markers. Instrument Preset value: All 9 markers Both commands are not available on the HP 3325A. Syntax Meaning 1 to 9 Table E-15. IMKS Response Format...
Page 264
Both commands are not available on the III’ 3325A. Syntax “index” I 1 to 9 1 Upto 9 separate markers are allowed. E-16. MMF Table “value” Range Given “units” Table E-17. IMMF Response Format Response Format Query Command I M M F E-18 HP 3324A Command Syntax...
Page 265
The MMS command selects whether multi-intervals or multi-markers is activated. Multi-Interval/ Marker Instrument Preset value: 1 (Multi-intervals) Both commands are not available on the HP 3325A. Meaning Multi-intervals selected, one marker/interval. Multi-markers selected, one interval. Table E-18. IMMS Response Format Query C ommand...
Page 266
(unmasked). They may be enabled or masked in any combination. Instrument Power-on,HP-IB Clear value: @ (no bits enabled) Instrument Preset value: not changed The MS command is available on the HP 3325h, but the IMS command is not. Syntax E-20 HP 3324A Command Syntax...
Page 267
Table E-19. Status Byte Mask Characters “character” System Fail Sweep Start Sweep Stop Program Em-o: O f f O f f O f f Table E-20. IMS Response Format Query Command Response Format HP 3324A Command Syntax E-21...
Page 268
The MT1 command is used to set the sweep time for the single interval when multi-markers are used. Sweep Time Instrument Power-on value: 1.0 sec. Both commands are not available on the HP 3325A. Syntax Table E-21. MTI “valuen Range Given “units” Description 100000 Seconds Table E-22.
Page 269
The MUP command is used to the single interval when multi-markers are used. Stop Frequency Instrument Power-on value: 10 MHz Both commands are not available on the HP 3325A. Syntax v a l u e Table E-23. MUP “value” Range Given “units” Range (sine) 0.0 to 21000000.000...
Page 270
The MUT command is used to set the start frequency MUT; Multi-Marker for the single interval when multi-markers are used. Start Frequency Instrument Power-on value: 1 MHz Both commands are not available on the HP 3325A. Syntax Table E-25. MUT “value” Range Given “units” “value” Range (sine) “units” Description 0.0 to 21000000.000...
Page 271
Chapter 5 under the heading “AC with DC Offset”. Instrument Preset value: 0.0 V(p-p) Both commands are available on the HP 3325A. Syntax v a l u e u n i t s Table E-27. OF 5ralue” Range Given “units”...
Page 272
OOF; Output The OOF command switches the main signal output on or off. Enable Instrument Preset value: 0 Both commands are not available on the HP 3325A. Syntax Meaning Disable main signal output. Enable main signal output. Table E-29. IOOF Response Format...
Page 273
PC; Phase The PC command causes a phase calibration between two HP 3324A’s, one of which has option 003 (slave) Calibration installed and the other option 004 (master). The PC command can only be sent to the instrument with option 003 installed.
Page 274
Both commands are available on the IIP 3325A. Syntax v a l u e 1 Table E-30. PH “value99 Range Given “units” “value” Range “units” Description -720.0 to 720.0 Degrees Table E-31. IPH Response Format Query Command Response Format E-28 HP 3324A Command Syntax...
Page 275
If either the SNR or SNI command is sent, all of the stores are cleared. This also happens with an instrument reset, and at power-on if the instrument is in the HP 3325A compatible mode. Command is available on the HP 3325A.
Page 276
OOF is the command used to control the main output signal, RF only determines where the signal will be present. 1 (front) Instrument Preset value: Both commands are available on the HP 3325A. Syntax Table E-32. IRF Response Format Query Comma4 Respons Format...
Page 277
3 ms logarithmic interval). Instrument Preset value: 0 (automatic) Both commands are not available on the HP 3325A. Syntax v a l u e Table E-33. RTT’ “value” Range Given “units” 100000 Seconds Table E-34.
Page 278
The SC command starts a continuous sweep. If the instrument is already sweeping, this command stops the Continuous Sweep sweep and does not restart it. FR can also be used to stop a sweep. Command is available on the HP 3325A. syntax E-32 HP 3324A Command Syntax...
Page 279
ISGS Intervals causes the instrument to output a list of the sequence. Both commands are not available on the HP 3324A. Syntax Meaning 1 to any of “SNI” The number of the interval.
Page 280
XSM and commands. IXSM Instrument Preset value: 1 Both commands are available on the HP 3325A. Syntax Table E-36. ISM Response Format Query Command Response Format E-34 HP 3324A Command Syntax...
Page 281
Note Sending the SNI command causes all of the stores to be cleared, and all of the intervals to be set to the default value. Both commands are not available on the HP 3325A. Syntax Meaning 1 to 50 Enable this number of intervals.
Page 282
Sending the SNR command causes all of the stores to be cleared, and all of the intervals to be set to the default value. Both commands are not available on the HP 3325~4. S y n t a x s p a c e...
Page 283
To set and read the stop frequency of other sweep intervals, use the XSP and IXSP commands. Instrument Preset value: 10 MHz Both commands are available on the HP 3325A. Syntax u n i t s v a l u e Table E-39.
Page 284
If either the SNR or SNI command is sent, all of the stores are cleared. This also happens with an instrument reset, and at power-on if the instrument is in the HP 3325A compatible mode. Command is available on the HP 3325A.
Page 285
Start Frequency. If the instrument is already in the sweep-reset state, this command starts a single sweep. If the instrument is sweeping, this command stops the sweep and does not restart it. Command is available on the HP 3325A. Syntax HP 3324A Command Syntax E-39...
Page 286
1 MHz Instrument Preset value: Syntax Table E-41. ST “value” Range Given “units” 21000000.000 0.0 to 0.0 to 21000.000000 0.0 to 21.000000 1 MH 1 Megahertz 1 Table E-42. IST Response Format Query Command Response Format E-40 HP 3324A Command Syntax...
Page 287
Cable Delay connecting the Phase Cal Out of the master (option 004) to the Phase Cal In of the slave (option 003). Instrument Preset value: 0 Command is not available on the HP 3325A. Syntax value un I ts Table E-43. TCD %aluen Range Given “units”...
Page 288
TE; Perform The TE command causes the instrument to perform a self-test. Self-test Command is available on the HP 3325A. syntax E-42 HP 3324A Command Syntax...
Page 289
XT1 and IXTI commands. Instrument Preset value: 1.0 sec. Both commands are not available on the HP 3325A. Syntax v a l u e Table E-45. TI “value” Range Given “units”...
Page 290
Cable Delay the device-under-test with the rear-signal outputs of the two HP 3324A’s containing options 003 and 004, when an automatic phase calibration is made. Instrument Preset value: 0 ps Command is not available on the HP 3325A.
Page 291
MF and IMF commands, respectively, when an index of 1 is used. Instrument Preset value: 5.0 MHz Both commands are not available on the HP 3325A. Syntax v a l u e u n i t s Meaning to any of “SNI”...
Page 292
1 is used. (Linear sweep mode) Instrument Preset value: 1 Both commands are not available on the HP 3325A. Syntax Meaning 1 to any of “SNI” Sets mode of interval. Selects Logarithmic sweep mode. Table E-51. IXSM Response Format...
Page 293
The XSP and IXSP commands are identical to the SP and ISP commands, respectively, when an index of 1 is used. 10 MHz Instrument Preset value: Both commands are not available on the HP 3325A. Syntax un its v a l u e index Meaning 1 to any of “SNI”...
Page 294
ST and IST commands, respectively, when an index of 1 is used. Instrument Preset value: 1 Both commands are not available on the HP 3325-4. Syntax 1 to any of “SNI” Sets start frequency of interval. Table E-54. XST Yralue” Range Given “units”...
Page 295
Instrument Preset value: 1.0 sec. are not available on the III’ 332511. S y n t a x I to auy of “SNI” Table E-56. XTI “value” Range Given “units” 0.0 to 100000 Seconds Table E-57. IXTI Response Format HP 3324A Command Syntax E-49...
Page 297
Error Messages Introduction This appendix contains a list of all the messages that can be obtained when the HP 3324A encounters an error. The following list contains all of the errors possible when operating the HP 3324A from the front panel.
Page 298
MESSAGE: Command: “HPIB addr.: xx” Cause: You are trying to change the HP-IB address, while the device is being addressed. Remedy: Un-address the HP 3324A via the controller before changing the HP-IB address. MESSAGE: Command: The frequency is either too small or too Cause: large for the current waveform.
Page 299
MESSAGE: HVamp/dBm incomp Command: “highvolt ON/off’ or (Ampl Cause: The amplitude cannot be in dBm for the high-voltage option. Enter the amplitude in Vrms or V(p-p). Remedy: index out range MESSAGE: “MARKER” “INTERVAL” Command: An index for a sweep interval has been Cause: entered which does not exist.
Page 300
log swp up only MESSAGE: Command: A log sweep has been programmed to be Cause: swept downwards. Re-enter the log sweep stop-frequency Remedy: to be higher than that of the start-frequency. no log swp <lHz MESSAGE: Command: A log sweep has been set with a Cause: start-frequency of <...
Page 301
For a DC and AC signal see the amplitude/offset table in chapter 5. MESSAGE: phase cal Wed Command: Cause: The attempted phase calibration has failed. Check to make sure that the two HP Remedy: 3324As are connected properly. MESSAGE: phase out range Command: (Phase) Cause: The phase value entered is not valid.
Page 302
MESSAGE: seq > 100 interv “INTERVAL” (m) Command: Cause: The sequence entered contains more than 100 intervals. Remedy: Reduce the number of intervals to 100 or less. MESSAGE: Command: [Reset Cause: The start frequency for the interval is too large. Remedy: Enter a smaller start frequency.
Page 303
MESSAGE: swp time < 10 ms Command: “SWEEP-TIME” Cause: The input sweep time is less than 10 ms. Remedy: Enter a sweep time for the interval of MESSAGE: swp time>lOOOOOs Command: “SWEEP-TIME” Cause: The input sweep time is greater than 100000 s.
Page 304
wrong seq index MESSAGE: Command: “INTERVAL” Cause: One of the indices in the sweep sequence does not exist. Remedy: Take the index out of the sweep sequence. F-8 Error Messages...
Page 305
The following lists the programming errors that can be obtained and evaluated using the IER program error query: Error Description Error Number Entry parameter out of bounds Invalid delimiter Frequency too high for waveform function Sweep time too small or too large Offset - amplitude incompatible Sweep frequency too large for function, start frequency too small,...
Page 306
The following lists the system errors that can be obtained and evaluated using the ISE system error query: System System Failure Description Failure Number Amplitude calibration failed Phase calibration failed External reference unlocked Main oscillator unlocked Self-test failed (report actual number to Service personnel).
Page 307
Backdating Introduction This appendix contains backdating information, which adapts this manual to instruments with serial numbers other than that shown on the title page. Changes are listed in the serial number order that they occured in the manufacture of the instrument. However, in adapting this manual to an instrument with serial number lower than that shown on the title page, apply the changes in reverse order.
Page 309
Bus management lines (HP-IB), D-4 Calibration amplitude, 5-8 Changing the bus address, 8-4, 9-11 Clear command, D-8 Clear lockout, D-9 Command groups, 9-7 Command syntax, E-l Compatibility mode setting the HP 3325A mode, 8-4 Compatibity mode HP 3325A compatible power-on state, 3-3 Connector Index-l...
Page 310
Data transfer modes, 9-7 DC offset, 5-10 high-voltage option, 5-13 no AC function, 5-11 with AC function, 5-11 Default address, 8-4, 9-11 Description of the HP-IB, D-l Display, 2-l Display indicators “compatib: yes/NO”, 8-4 2-5, 3-6 “HIGH-VOLT”, “highvolt: on/OFF”, 8-7...
Page 312
frequency, function generator, modify mode, 3-13, 5-4 programming as a function generator, 1 I-10 programming as a multi-interval sweep generat)or, 1 l-l I programming as a multi-marker sweep generator, 11-13 recalling an instrument state, 3-16 saving an instrument state, 3-16 serial poll, 10-9 setting up a multi-interval sweep, 7-l 1 setting up a multi-marker sweep, 7-20...
Page 313
9-4 HP 3324A capabilities, 9-l logic levels, C-7 messages, 9-6 setting the address, 8-4, 9-11 Impedance of high-voltage output, 2-3, 2-4, A-11 of main signal output, 2-3, 2-4 Input signals, 3-4 Instrument introduction, l-l Interpreter, 9-2 Intervals setting number of, 8-3...
Page 314
7-22 3-11 3-11 2-1, 7-l 2-1, 7-l Linear sweep interval, 7-4 Local lockout, D-9 Local mode, l-2 Logarithmic sweep interval, 7-4 Main signal output connector, 2-3, 2-4, 3-4 Marker connector, 2-5 Marker frequency m u l t i - i n t e r v a l 7 - 5 , 7 - 1 5 sweep, Marker output connector, 3-7 Messages...
Page 316
Setting the bus address, 8-4, 9-11 Setting up a multi-interval sequence, 7-8 Setting up a multi-marker interval, 7-17 Shift key, 3-9 Signals input/output, Single sweep, 7-7, 7-17 Software revision, 8-7 Specifications, A-l forcing an, 10-6 State sweep default, 7-2 turn-on, 3-2 Status byte, 10-l Status reporting, 10-l Stores...
Page 317
Talk/listen addresses, D-6 Tests self-test, 8-5 Trigger (HP-IB), D-10 Turn-on state, 3-2 Utilities, Viewing the bus address, 8-4, Y-11 Voltage AC line, C-3 Warm-up time, A-2 Waveforms, 6- 1 X-drive connector, 2-5, 3-8 Z-blank connector, 2-5, 3-7 Index-9...
Page 319
Sales and Service Office Directory Information Hewlett-Packard products are sold and supported worldwide through HP offices. To contact the closest HP Sales and Service Office, please check your telephone directory, or contact one of the HP Headquarters listed below. Asia Hewlett-Packard Asia Ltd.
Page 320
Eastern Europe Lieblgasse 1 P.O. Box 72 1222 Vienna Austria Telephone: (222) 2500-O Telex: 13 4425 HEPA A Hewlett-Packard S.A. Northern Europe V.D. Hooplaan 241 P.O. Box 999 118 LN 15 Amstelveen Netherlands Telephone: 20 547 9999 Telex: 189 19 hpner South East Europe Hewlet-Packard S.A.
Page 321
Middle East and Hewlett-Packard S.A. Middle East/Central Africa Sales H.Q. Central Africa 7, Rue du Bois-du-Lan P.O. Box 364 1217 Meyrin 1 Geneva Switzerland Telephone: (022) 83 12 12 Telex: 27835 hmea Cable: HEWPACKSA Geneve United Kingdom Hewlett-Packard Ltd. Nine Mile Ride Wokingham Berkshire, RGll 3LL Telephone: 0344 773 100...
Page 322
Midwestern USA Hewlett-Packard 5201 Tollview Drive Rolling Meadows IL 60008 Telephone (312) 255-9800 Southern USA Hewlett-Packard 2000 South Park Place Atlanta CA 30339 Telephone: (404) 955-1500 Western USA Hewlett-Packard 5161 Landershim Roulevard North Hollywood CA 91601 Telephone: (818) 505-5600 Hewlett-Packard Other International Intercontinental Operations...